PIROP

PIROP is a modern ultrasound corneal pachymeter with biometry function, using for the first time the ultra high 400 MHz sampling frequency, giving the users new idea of superb accuracy and repeatability.

Key advantages:

  • Full measurement diagnostics of all kinds of eye
  • Very high accuracy and repeatability of measurement

PACHYMETRY

Pachymetry Probe
Styles Straight Pachymeter Probe
Frequency
20 MHz ± 10%
Probe Tip Diameter
1,9 mm       
0,075”
Acceptance Angle
10º
Method
contact
Weight
21 g       
0,046 lbs
Cable length with probe
1650 mm   
65”
Corneal thickness measurements


Sampling frequency
400 MHz
Measurement Accuracy
±2 microns
Measurement Precision ±1 micron
Measurement Resolution
±1 micron
Operational Modes
AUTO, MANUAL
Measurement range
0,25 – 1,1 mm   
for default velocity: 1640 m/s
Methods
central measurement or cartographic map (automatic continuous scanning)
Cartographic maps
1
5, 9, 13, 21, 25        
1 MULTI           
5 MULTI, 9 MULTI              

- Single Reading / One Point
- Single Reading / Multiple Points
- Multiple Readings / One Point   
- Multiple Readings / Multiple Points   
Intra-Ocular Pressure Correction (IOP)
- four predefined formulas: Kohlhaas and Shah + Herndon and Doughty + Whitacre + Ehlers
- 9 user defined formulas
- cornea thickness value for IOP:
   - from measurement
   - entered manually
Statistical Analysis Average of 512 Individual Measures for Each Reading
Average of up to 12 Readings, for 1 MULTI map
Average of up to 5 Readings for Each Point, for 5 MULTI and 9 MULTI maps
Average Across All Points
Standard Deviation of Up to 12 Readings, for 1 MULTI map
Standard Deviation of Up to 5 Readings for Each Point, for 5 MULTI and 9 MULTI maps
Corneal Velocity
1400 ÷ 2000 m/s
default value: 1640 m/s
DELAY of measurement
1 ÷ 9 sec
default value: 2 sec
Bias correction
60 ÷ 130 % default value: 70%
Automatic Calibration Check
   
Built-in Probe Sensitivity Test    


BIOMETRY

A-scan probe

Transducer Frequency
12 MHz
Axial resolution
120 µm
Diameter of tip
7 mm        
0,275”
Method
contact, immersion
Aiming beam
LED red pointer
Points on X
512
Levels on Y
256
Depth of exploration
40 mm       
1,57”
Weight
34 g       
0,075 lbs
Cable length
1500 mm   
59”
Axial length measurements

Types of measurement
anterior chamber, lens, vitreous, total length,
Type of eye
Phakic, Cataract, Dense Cataract, Aphakic, Pse/Acrylic, Pse/PMMA, Pse/silicon,
Measurement mode
MANUAL 1, AUTO 1   
MANUAL 2, AUTO 2   
AC, LENS, VITR, AXL
AC, AXL
Amplifier
Gain 70 – 100%,  automatic TGC control, digital filter
Clinical resolution
± 0,1mm (±0,004”)
Velocity
adjustable for each segment
Calibration
semiautomatic, with calibration cylinder
Measurements
results for 10 scans for anterior chamber depth (AC) lens thickness (LENS), vitreous length (VITR) axial length (AXL), the results can be printed and stored in USB memory.
Statistics
average and standard deviation.
Frozen image
foot pedal or push button
IOL formulas
SRK II, SRK T, Holladay, Hoffer-Q, Binkhorst II, Haigis
IOL Storage Capacity
10 predefined types stored


GENERAL FEATURES

Display

Type LCD 5,7” Color Display with Touch Screen Overlay
Resolution
640 x 480 pixels (h x v)
Width
116 mm
4,5”
Height
87 mm
3,5”
General software features

  • Inbuilt calendar and clock
  • 9 Customizable User Profiles
  • Reports archiving on USB pendrive
  • Reports printout
  • Easy software updating
External printer

Interface USB
Paper Size
- width           
- roll diameter       
57 mm        2,25”
33 mm        1,3”
Electrical requirements

External power supply 100÷240V, 50/60Hz to 12V DC
Power consumption
10 W
IEC protection class
II
Peripherals

DC input pin Jack Connector
Foot Pedal
pin Jack Connector
Probe
mini coaxial
Printer
USB connector
External memory
USB connector
Weight

Weight of unit 0,84 kg       
1,85 lbs
with complete equipment
2,46 kg       
5,4 lbs
Overall dimensions

Width
199 mm       
7,8”
Depth
154 mm       
6”
Height
84 mm       
3,3”
Standard equipment

  • PIROP Unit
  • Foot Pedal
  • AC Adapter
  • Pachymeter probe
  • Printer with paper
  • USB cable for printer
  • Carrying case
Standards and norms

  • Medical Device Directive 93/42 EEC,
  • EMC Directive 89/336/EEC,
  • Electromagnetic Compatibility EN 60601-1-2,
  • Electrical Safety IEC 601