PIROP
PIROP is a modern ultrasound corneal pachymeter with biometry function, using for the first time the ultra high 400 MHz sampling frequency, giving the users new idea of superb accuracy and repeatability.
Key advantages:
- Full measurement diagnostics of all kinds of eye
- Very high accuracy and repeatability of measurement
PACHYMETRY
| Pachymetry Probe |
| Styles |
Straight Pachymeter Probe |
Frequency |
20 MHz ± 10% |
|
Probe Tip Diameter |
1,9 mm
|
0,075” |
Acceptance Angle |
10º |
|
Method |
contact |
|
Weight
|
21 g
|
0,046 lbs |
Cable length with probe
|
1650 mm
|
65” |
Corneal thickness measurements
|
|
|
Sampling frequency
|
400 MHz |
|
Measurement Accuracy |
±2 microns |
|
| Measurement Precision |
±1 micron |
|
Measurement Resolution |
±1 micron |
|
Operational Modes |
AUTO, MANUAL |
|
Measurement range |
0,25 – 1,1 mm
|
for default velocity: 1640 m/s |
Methods |
central measurement or cartographic map (automatic continuous scanning) |
Cartographic maps |
1 5, 9, 13, 21, 25 1 MULTI 5 MULTI, 9 MULTI
|
- Single Reading / One Point - Single Reading / Multiple Points - Multiple Readings / One Point - Multiple Readings / Multiple Points
|
Intra-Ocular Pressure Correction (IOP) |
- four predefined formulas: Kohlhaas and Shah + Herndon and Doughty + Whitacre + Ehlers - 9 user defined formulas - cornea thickness value for IOP: - from measurement - entered manually |
| Statistical Analysis |
Average of 512 Individual Measures for Each Reading Average of up to 12 Readings, for 1 MULTI map Average of up to 5 Readings for Each Point, for 5 MULTI and 9 MULTI maps Average Across All Points Standard Deviation of Up to 12 Readings, for 1 MULTI map Standard Deviation of Up to 5 Readings for Each Point, for 5 MULTI and 9 MULTI maps |
Corneal Velocity |
1400 ÷ 2000 m/s
|
default value: 1640 m/s
|
DELAY of measurement
|
1 ÷ 9 sec
|
default value: 2 sec |
Bias correction
|
60 ÷ 130 % |
default value: 70% |
Automatic Calibration Check
|
|
|
| Built-in Probe Sensitivity Test |
|
|
BIOMETRY
| A-scan probe |
|
|
Transducer Frequency
|
12 MHz |
|
Axial resolution |
120 µm |
|
Diameter of tip |
7 mm
|
0,275” |
Method |
contact, immersion |
|
Aiming beam |
LED red pointer |
|
Points on X |
512 |
|
Levels on Y |
256 |
|
Depth of exploration |
40 mm
|
1,57” |
Weight
|
34 g
|
0,075 lbs |
Cable length
|
1500 mm
|
59” |
| Axial length measurements |
|
|
Types of measurement
|
anterior chamber, lens, vitreous, total length, |
Type of eye |
Phakic, Cataract, Dense Cataract, Aphakic, Pse/Acrylic, Pse/PMMA, Pse/silicon, |
Measurement mode |
MANUAL 1, AUTO 1 MANUAL 2, AUTO 2 |
AC, LENS, VITR, AXL AC, AXL
|
Amplifier |
Gain 70 – 100%, automatic TGC control, digital filter |
Clinical resolution |
± 0,1mm (±0,004”) |
|
Velocity
|
adjustable for each segment |
|
Calibration
|
semiautomatic, with calibration cylinder |
|
Measurements |
results for 10 scans for anterior chamber depth (AC)
lens thickness (LENS), vitreous length (VITR) axial length (AXL), the
results can be printed and stored in USB memory. |
Statistics |
average and standard deviation. |
|
Frozen image |
foot pedal or push button |
|
IOL formulas |
SRK II, SRK T, Holladay, Hoffer-Q, Binkhorst II, Haigis |
IOL Storage Capacity
|
10 predefined types stored |
|
GENERAL FEATURES
| Display |
|
|
| Type |
LCD 5,7” Color Display with Touch Screen Overlay |
Resolution |
640 x 480 pixels (h x v) |
Width
|
116 mm
|
4,5” |
Height
|
87 mm
|
3,5” |
| General software features |
|
|
- Inbuilt calendar and clock
- 9 Customizable User Profiles
- Reports archiving on USB pendrive
- Reports printout
- Easy software updating
|
| External printer |
|
|
| Interface |
USB |
|
Paper Size
|
- width - roll diameter
|
57 mm 2,25” 33 mm 1,3” |
| Electrical requirements |
|
|
| External power supply |
100÷240V, 50/60Hz to 12V DC |
Power consumption
|
10 W |
|
IEC protection class
|
II |
|
| Peripherals |
|
|
| DC input |
pin Jack Connector |
|
Foot Pedal |
pin Jack Connector |
|
Probe |
mini coaxial |
|
Printer
|
USB connector |
|
External memory
|
USB connector |
|
| Weight |
|
|
| Weight of unit |
0,84 kg
|
1,85 lbs |
with complete equipment
|
2,46 kg
|
5,4 lbs |
| Overall dimensions |
|
|
Width
|
199 mm
|
7,8” |
Depth
|
154 mm
|
6” |
Height
|
84 mm
|
3,3” |
| Standard equipment |
|
|
- PIROP Unit
- Foot Pedal
- AC Adapter
- Pachymeter probe
- Printer with paper
- USB cable for printer
- Carrying case
|
| Standards and norms |
|
|
- Medical Device Directive 93/42 EEC,
- EMC Directive 89/336/EEC,
- Electromagnetic Compatibility EN 60601-1-2,
- Electrical Safety IEC 601
|