DESMIN
A new generation of ophthalmic ultrasound scanners Desmin enables the estimation of physiological and pathological structures of the eyeball and orbit, full automatic and manual biometry and calculation of lens power for intraocular implants. The system works in A, B, and pachymetry modes. Desmin utilizes the newest technologies in electronics and gives to a doctor a powerful and reliable tool.
SPECIFICATION B-SCAN MODE
| B-scan probe: |
| Scan type |
mechanical convex |
| Transducer Frequency |
12 MHz |
Diameter of tip |
22 mm
|
0,867” |
Focus |
20 - 24 mm
|
0,79” - 0,94” |
Depth of exploration |
20, 30, 40, 60 mm
|
0,79” - 2,37” |
Frame |
55°, 256 lines, 512 samples per line |
Frame rate |
11 frames/sec |
Axial resolution |
120 µm |
Lateral resolution |
300 µm |
Weight
|
165 g
|
0,364 lbs |
Cable length
|
1500 mm
|
59” |
| Amplifier: |
| Gain |
adjustable from 63 dB to 105 dB |
TGC
|
adjustable from 0 dB to 30 dB, five zones |
Gain curves
|
linear, logarithmic |
Display:
|
Display mode
|
B, B+A, B+B, A, B ZOOM (1,2x – 2,2x) |
Image |
B-scan with simultaneous selectable vector A-scan |
Image orientation |
up/down |
4 images memory |
2 for right and 2 for left eye |
Gray Scale Levels |
256 |
| Image and data storage |
- images with patient’s data and examination parameters stored in the internal flash memory (option), up to 1000 images can be stored - USB external memory - external PC with frame grabber
|
Processing
|
smooth image, color coded image generation, brightness, contrast |
| Post processing Tools: |
| Measurements |
- three sets of distance calipers with variable velocity: for anterior, lens, and posterior measurement - two sets of angle calipers with predefined velocity - circumference, area and volume measurements - text annotations - measurement report |
Grey scale curves
|
Logarithmic, Exponential, Linear, S |
Image |
dark inverse, bright inverse, positive, zoom 1,2x – 2,2x up and down image orientation |
Dynamic
reply
|
Cineloop movie sequence up to 23 seconds, max 256 images, 11
images a second, real time viewing or frame by frame |
BIOMETRY
| A-scan probe: |
| Transducer Frequency |
12 MHz |
|
Axial resolution |
120 µm |
|
Diameter of tip |
7 mm
|
0,275” |
Method |
contact, immersion |
|
Aiming beam |
LED red pointer |
|
Points on X |
512 |
|
Levels on Y |
256 |
|
Depth of exploration |
30, 40, 60 mm
|
1,18” - 2,37” |
Weight
|
34 g
|
0,075 lbs |
Cable length
|
1500 mm |
59” |
| Axial length measurements: |
| Types of measurement |
anterior chamber, lens, vitreous, total length, |
Type of eye |
Phakic, Cataract, Dense Cataract, Aphakic, Pse/Acrylic, Pse/PMMA, Pse/silicon, |
Measurement mode |
automatic (precise, fast, advanced algorithm), manual, automatic detection of scleral spike for advanced algorithm |
Amplifier |
Gain 63 dB to 105 dB, automatic TGC control, digital filter |
Color mode |
white, yellow, green |
Frozen image |
auto, auto + save, manual (foot pedal) |
Manual correction |
adjustable markers; for retina, posterior lens, anterior lens spike |
Velocity |
adjustable for each segment |
Echo detection level |
adjustable from 60 to 100% (in 10% steps) |
Calibration |
semiautomatic, with calibration cylinder |
Accuracy
|
Electronic ± 0,0586mm (±0,0023”), Clinical ± 0,1mm (±0,004”) |
Measurements memory |
results for 10 scans for anterior chamber depth
(ACD) lens thickness (LENS), vitreous length (V.) axial length (AL) |
Report Data |
Patients Name, P.Id number, Eye examined, K-readings,
User Name, Date, Time, Immersion On/OFF, measurement results, IOL
calculations |
Statistics |
average and standard deviation for ACD, LENS, V, AL |
IOL formulas
|
SRK II, SRK T, Holladay, Hoffer-Q, Binkhorst, Haigis, AvgIOL, |
IOL Storage Capacity |
10 measurements |
Calculations
|
two tables of various IOL calculations 6 various implant powers for desired ametropia value implant values available in increment of: 0,50D |
PACHYMETRY
| Pachymetry Probe: |
| Styles |
Straight Pachymeter Probe |
Frequency |
20 MHz ± 10% |
|
Probe Tip Diameter |
1,9 mm
|
0,075” |
Acceptance Angle |
10º |
|
Method |
contact |
|
| Weight |
21 g
|
0,046 lbs |
Cable length
|
1500 mm
|
59” |
Corneal thickness measurements:
|
Measurement range
|
0,3 ÷ 1,3 mm
|
for default velocity: 1640 m/s |
Measurement resolution |
±1 micron |
|
Measurement precision |
±1 micron |
|
Operational Modes
|
auto, manual |
|
Corneal Velocity |
1200 ÷ 2400 m/s
|
default value: 1640 m/s |
Delay of measurement |
1 ÷ 30 sec
|
|
Bias correction
|
50 ÷ 150%, |
|
Methods |
central measurement or cartographic map (automatic continuous scanning) |
Cartographic maps |
1 5, 13, 21 |
- Single Reading / One Point - Single Reading / Multiple Points
|
Intra-Ocular Pressure Correction (IOP) |
- two predefined formulas: Shah and Kohlhaas + Ehlers - one user defined formula - value for IOP correction entered manually from tonometer |
| Statistical Analysis |
actual value - average of 512 individual measures for each reading Average Across All Points, Minimum, Maximum Value |
| General features: |
- LCD: 12” 1024 x 768 color screen SVGA
- Integrated intuitive silicone keyboard with trackball, backlighted
- USB 2GB pendrive external memory
- Reports archived on USB pendrive or printed out on videoprinter
- Automatic probe shut off after 10 minutes of continuous running (setting from 3 to 30 minutes)
- ESR (Enhanced Speckles Reduction) – reduction of granularity of image
- SAS (Spectral Adaptive System) – automatic resolution adjustment
- Inbuilt clock (date and time)
- Foot pedal
- 5 Customizable User Profiles
|
| Electrical requirements: |
| Power supply |
100/240 V, 50/60 Hz |
Power consumption
|
30 VA |
IEC protection class
|
I |
protection level type B, probes type BF |
Peripherals:
|
|
|
| A probe |
pin connector |
|
B probe |
pin connector |
|
Pachymerty |
mini coaxial |
|
Foot Pedal |
jack connector |
|
Printer |
coaxial, jack connector |
External memory
|
USB connector |
|
External monitor
|
coaxial |
|
| Optional equipment: |
|
|
- Optional equipment
- Trolley
- Videoprinter
- Internal memory module
- Frame grabber PC connection
|
| Overall dimensions: |
|
|
| Width |
310 mm
|
12,2” |
Depth |
155 mm
|
6,1” |
Height
|
290 mm
|
11,5” |
Weight
|
5 kg
|
11 lbs |
| Standards and norms: |
|
|
- Medical Device Directive 93/42 EEC
- EMC Directive 89/336/EEC
- Emission - Class A
- Compatibility EN 61000
- Electrical Safety IEC 601
|